(Cat No. - 9178) |
Three level safety ciruits:
Constant indication and control
of air & skin temperature.
Individual sensors for air & skin.
L.E.D. bar graph with 4 increments, to
indicate heater output
Servo Specification:
A1 Over Temp. cut off |
37.5oC |
A2
Compenent failure |
0.7oC, above
set value automatically |
A3
System fail |
38oC parallel electrical
cutoff. |
Max. Temp.
setting |
380C |
Alarms:
Over
temperature |
Audio Visual |
Sensor fail |
Audio Visual |
Check Ambient |
Audio Visual |
System fail |
Audio Visual |
Technical
Specifications :
Model |
AD 18 |
Length |
27 inch. (685 mm) |
Depth |
17 inch. (711 mm) |
Height |
51 inch. (1700 mm) |
Weight |
40 kgs approx. |
Baby bed |
Fabricated in stainless steel |
Demensions |
28 x 18(711 x 457
mm) |
Tilt |
-8 Degrees |
Storage |
2 open
compartments in stainless steel |
Working level |
Approx. 950mm from
ground |
Best environment
for operation |
220 C to 290 C |
Electrical Rating |
230 V.A.C. - 10% 50 Hz |
Heater wattage |
Maximum 500
Watts |
Observation
lamp |
60 Watts |
Sensor |
Flat skin
probe |
Heater |
2 Silicon quartz crystal heaters |
Electronic Controller |
3 digit green L.E.D. display for patients temp. |
Display resolution |
0.10 C |
|